Operation – Leica Biosystems CV5030 User Manual

Page 53

Advertising
background image

53

Leica CV5030 – Robotic Coverslipper

5.7 Display indicators and instructions (continued)

5. Operation

Display

Explanation

Action

At least three cover slips in a

row are defective.

Check the cover slips in the

cover slip magazine.

If the messages occurs repeat-

edly despite the cover slips be-

ing intact, clean the cover slip

sensor. If the message persists,

clean the Pick & Place module

(skids and suction cups).

The user has pressed the STOP

button.

The output magazine, if loaded,

has to be removed from the

instrument. Grabbed specimen

slides or held cover slips also

must be removed from their

position by pressing the respec-

tive RELEASE SLIDE or RELEASE

COVERSLIP button.

Pressing START reinitializes the

instrument. Afterwards, proceed

the same as when switching

on the instrument (

refer to

Chapter 5.3.

).

Important!

If the instrument is operated as

a workstation, the user must en-

sure that before reinitialization,

no rack is in the transfer station.

The coverslipping operation

was ended. The number after

FINISHED corresponds to the

selected SET 1-4 program.

Remove the cover slipped speci-

men slides.