Malfunctions and troubleshooting, 1 error codes (continued) – Leica Biosystems CV5030 User Manual

Page 90

Advertising
background image

90

Instructions for Use V 3.0 RevE – 04/2013

8.

Malfunctions and Troubleshooting

8.1 Error codes (continued)

Display

Cause

Troubleshooting

Error 313

TS-Z BLOCKED

The transfer arm of the

TS5025/5015 Transfer Sta-

tion is blocked in its up/

down movement (Z-axis).

Remove the cause of the blockage (specimen slide holder) and

inspect the transfer arm of the TS5025/TS5015. Remove empty

specimen slide holders from the discharge chute of the transfer

station. Switch the instrument off and on. Proceed further as de-

scribed in

Chapter 5.3

. Place any removed specimen slide holders

into the coverslipper and have them coverslipped (thus outside of

the workstation operation).

Error 314

TS-AX BLOCKED

The gripper of the TS5015

Transfer Station is blocked

in its left/right movement

(X-axis).

Remove the cause of the blockage (specimen slide holder) and

inspect the gripper of the TS5015. Remove empty specimen

slide holders from the discharge chute of the transfer station.

Switch the instrument off and on. Proceed further as described

in

Chapter 5.3

. Place any removed specimen slide holders into

the coverslipper and have them coverslipped (thus outside of the

workstation operation).

Error 315

TS-AY BLOCKED

The carriage of the TS5015

Transfer Station is blocked

in its forward/backward

movement (Y-axis).

Remove the cause of the blockage (specimen slide holder) and

inspect the carriage of the TS5015. Remove empty specimen

slide holders from the discharge chute of the transfer station.

Switch the instrument off and on. Proceed further as described

in

Chapter 5.3

. Place any removed specimen slide holders into

the coverslipper and have them coverslipped (thus outside of the

workstation operation).

Error 316

TS-AZ BLOCKED

The gripper of the T5015

Transfer Station is blocked

in its up/down movement

(Z-axis).

Remove the cause of the blockage (specimen slide holder) and

inspect the gripper of the TS5015. Remove empty specimen

slide holders from the discharge chute of the transfer station.

Switch the instrument off and on. Proceed further as described

in

Chapter 5.3

. Place any removed specimen slide holders into

the coverslipper and have them coverslipped (thus outside of the

workstation operation).

Error 319

CS SENSOR DEF.

The cover slip sensor is

stuck or defective.

Clean the Pick & Place module and the cover slip sensor using

a lint-free cloth soaked in compatible solvent as described in

Chapter 7.5.10

.

Error 322

CONFIG FAULT

The reference values of the

instrument are faulty.

Contact Leica Service.