Leica CV5030 – Robotic Coverslipper

6. Workstation operation

• All empty specimen slide holders must be

removed from the discharge chute and the

drawer of the transfer station must be closed.

Then, press the START button so that the

"READY" message appears on the display.

Transfer from the slide stainer to the robotic

coverslipper is now possible once again.

6.1 Operation as ST5010 – CV5030 Workstation (continued):

Basically, we do not recommend man-

ual loading of the CV5030 in worksta-

tion operation, as this can impair or

stop the program flow of the ST5010.

So that you do not disturb workstation

operation, you should process speci-

men slide holders that come from the

slide stainer first. Only then should you

process the specimen slide holders to

be loaded manually in the coverslipper.

• However, if it is necessary to insert a specimen

slide holder into the loading bath manually

for coverslipping, the empty specimen slide

holder must also be removed from the loading

drawer manually. The empty specimen slide

holder is not conveyed automatically into the

transfer station and thus onto the discharge

chute. Specimen slide holders from the stainer

have priority for processing.

• The display shows the "FINISHED" message

and a signal tone sounds. Confirm the signal

tone with the RESPOND button. This is fol-

lowed by the "PAUSED" and "CHECK BATH"

messages and a signal tone. Confirm the signal

tone with the RESPOND button. The display

then shows the "PAUSED" message.

Ensure that at this time, no specimen

slide holder is manually inserted into

the loading bath, as this can stop the

program flow.

• Do not press the START button until after the

specimen slide holder is removed. This will

ensure that pending specimen slide holders

from the slide stainer can be processed.

• If a specimen slide holder is waiting in the

unload station of the stainer to be transferred

to the CV5030, a manually loaded specimen

slide holder cannot be processed.

Manual loading