39

Leica CV5030 – Robotic Coverslipper

5.3 Switching the instrument on or off

The instrument initializes after being switched on. The installed

software version is shown in the display.
The "READY" LED is illuminated in red.

"CHECK BATH" is the prompt for the user to insert the loading

bath after initialization.

To do so, open the loading door and pull out the loading draw-

er. This is the last chance to remove the cover from the bath.

5. Operation

LED

When the loading drawer with loading bath is pulled out,

"BATH NOT READY" appears on the display.

ð

Move the power switch on the front side of the instrument to

the ON = I or OFF = 0 position.
This is followed by a beep.

The digits on the display show the selected cover-

slipping program (

refer to Chapter 5.9

)

Important!

BEFORE switching off the instrument, remove the

loading bath, cover it and place it under a labora-

tory fume hood until the next time the instrument is

started. Only reinsert the loading bath AFTER the in-

strument is switched back on and initialized to avoid

changes to settings and damage to the gripper.