5.8 Button functions for programming

Fig. 32

Press and hold

for 2 s

Press 1 x

Fig. 33

71

72

5. Operation

Press and hold RESPOND for approx. two seconds to set

the programs and parameter sets the instrument uses to

operate.

The individual parameter sets are set using two sub-

menus:

Menu A (

Fig. 32

) and Menu B (

Fig. 33

)

Toggle between the two submenus by briefly pressing

RESPOND. The active submenu (71) is displayed in the

right half of the top display line.

There are four selectable programs (SET 1 through 4)

that are displayed on the upper left half of the display

(

Fig. 32,

72).

The bottom line in the display shows the parameters in

MENU A:
STK - Length of the coverslip mountant stroke
VOL - Quantity of the coverslip mountant
TYP - Pressure in the mountant bottle
LEN - Length correction for the coverslip mountant stroke

MENU B includes the following parameters:

CSP - Cover slip placement position
STP - Starting point for the coverslip mountant stroke
SPK - Volume of the signal tone
LIM - Cover slip limit setting