4 cleaning and maintenance as necessary, 2 dispenser needle cleaner (nozzle cleaner), Cleaning and maintenance as necessary – Leica Biosystems CV5030 User Manual

Page 82: Cleaning and maintenance

Advertising
background image

82

Instructions for Use V 3.0 RevE – 04/2013

7.

Cleaning and Maintenance

A

Filling the coverslipping mountant bottle:
• If possible, top up the coverslip mountant at the end of the working day so that

any air bubbles that arise during filling are degassed by the next working day.
• The coverslip mountant needs between 6-12 hours of resting time to degas (time

depends on the type of the coverslip mountant).
• If necessary, clean the bottle neck and both O-rings (blue = bottle neck and

black = dispenser group) to remove coverslip mountant residue.

B

Replace the active carbon filter if necessary, but no later than after three months (

refer

to Chapter 7.7.1

).

C

Check the transfer station for dirt and clean it if necessary:
• TS5015: Transfer arm (

refer to Chapter 7.7.3

) and carriage in y-direction.

• TS5025: Discharge chute (

refer to Chapter 7.7.2

) and transfer arm.

7.4 Cleaning and maintenance as necessary

7.5 Description of the required daily cleaning measures

7.5.1 Loading chute and bath transport with transport chain

• Check the loading chute and bath transport with transport chain for broken glass and adhe-

sive residue and carefully remove any dirt or debris.

• To clean adhesive residue, use a lint-free cloth soaked in compatible solvent. Loose glass

shards and splinters can be cleaned up using a commercially available vacuum cleaner.

7.5.2 Dispenser needle cleaner (nozzle cleaner)

• Fill the dispenser needle cleaner with approx. 5 ml of solvent.

• Check the brush. If the brush is hard or very dirty, replace it with a new brush.

There is a danger of cutting injuries during this cleaning step. Therefore, proceed with the

necessary caution.

Wear protective clothing!

Advertising