5.7 Display indicators and instructions (continued)

5. Operation

Display

Explanation

Action

The instrument is in pause mode.

The coverslipping process was

interrupted due to the empty

cover slip magazine.

Fill the cover slip magazine.

Pressing START allows work to

continue.

The instrument is in pause mode.

There is no loading bath in the

loading drawer.

Insert a loading bath into the

loading drawer. Press RESPOND

as confirmation, then press

START to continue coverslip-

ping.

The instrument is in pause mode

since a vacuum could not be

attained in the system.

Check the coverslip mountant

bottle/dispenser system for

leaks and whether both O-rings

are present; replace the O-rings,

if necessary.

The instrument is in pause mode;

the working pressure in the sys-

tem could not be attained.

Check the coverslip mountant

bottle/dispenser system for

leaks and whether both O-rings

are present; replace the O-rings,

if necessary.

The instrument is in pause mode;

the system was unable to grab a

specimen slide.

In the specimen slide holder,

check whether the specimen

slides are inserted correctly

and undamaged. If the message

occurs repeatedly despite the

specimen slides being inserted

correctly, remove the specimen

slides and apply cover slips by

hand.