7.

Cleaning and Maintenance

7.6.3 Specimen slide holder, gripper and

output magazines

• Carefully clean the gripper using a lint-free

cloth soaked in compatible solvent (if neces-

sary, also daily if using labels that protrude or

are sensitive to solvent).

• Check the output magazines for clinging and

dried coverslipping mountant.

• In particular, dried adhesive residue in the

slots of the output magazines can cause prob-

lems when inserting the specimen slides.

• Carefully clean the output magazines using a

lint-free cloth soaked in compatible solvent.

• Check the brush of the dispenser needle

cleaner for dried-on, hardened coverslipping

mountant.

• To do so, take the brush out of the plastic

container and remove clinging coverslip

mountant.

• If the brush is very dirty or hard, replace it with

a new one.

• Before beginning work, fill the plastic con-

tainer with 5 ml of solvent. To do so, use the

plastic pipettes provided.

Fig. 48

7.6.2 Dispenser needle cleaner

(nozzle cleaner)

7.6 Description of the required weekly cleaning measures (continued)

Notice: The output magazines must not

be placed in solvent over a period of

multiple hours (e.g. overnight) (

refer to

Chapter 7.1

).