1 dispenser group, Cleaning and maintenance – Leica Biosystems CV5030 User Manual

Page 85

Advertising
background image

85

Leica CV5030 – Robotic Coverslipper

7. Cleaning and Maintenance

7.6 Description of the required weekly cleaning measures

7.6.1 Dispenser group

We recommend rinsing the dispenser group once

a week using approx. 150 ml of a solvent that is

compatible with the coverslip mountant.

• Wipe the coverslip mountant off of the hose

end of the dispenser group.

• Insert the replacement bottle filled with solvent

into the instrument, screw on the lid to close it

correctly (ensure that both O-rings are seated

correctly), then switch the instrument back on.

• Switch off the instrument.
• Fill the additional glass bottle included in the

standard scope of delivery with 150 ml of sol-

vent.

• If a third glass bottle is needed because the

replacement bottle is already filled with cov-

erslip mountant, you can order an additional

glass bottle under order No.: 14 0464 36537.

• Unscrew the coverslip mountant bottle and

remove it from the instrument.

• Wait for the complete initialization to finish,

then proceed as for an instrument restart (

re-

fer to Chapter 5.3.

). However, the priming cycle

should be carried out until all of the solvent

has been flushed through the dispenser group

(press and hold the PRIME button).

Filling the coverslipping mountant bottle:
• If possible, at the end of the working day so

that any air bubbles that arise during filling are

dissipated by the next working day.

• The coverslip mountant needs between

6-12 hours of resting time to degas (time de-

pends on the type of the coverslip mountant).

• If necessary, clean the bottle neck and both

O-rings (blue = bottle neck and black = dis-

penser group) to remove coverslip mountant

residue.

Notice: Due to the high pressure when

rinsing with solvent, it is necessary to

wear gloves, safety goggles and ap-

propriate protective clothing when

changing the cover slip mountant! Se-

lect an adequately sized container to

catch the rinsing residue.

Notice: Danger of dripping – any

mountant that drips down must be

picked up immediately using an absor-

bent cloth.

Advertising