4.

Instrument Setup

• Fill the loading bath (

Fig. 27

, 68) with a solvent

compatible with the coverslip mountant. The

level should reach up to the specimen slide's

label field.

• Fill the glass vial (69) in the rest position with a

solvent compatible with the coverslip mountant

(approx. 10 ml). The dispenser needle must al-

ways be immersed in the solvent.

• The standard scope of delivery includes two

different cover slip magazines with different

widths (22/24 mm). Fill and insert the magazine

corresponding to the cover slip width being used

(possible cover slip lengths 40 - 60 mm) (

refer to

Chapter 9, "Optional Accessories"

).

Fig. 28

4.11 Refilling consumables

44

69

40

62

In order to be able to degas air

bubbles generated while filling the

mountant bottle (44), the coverslip

mountant must sit for 6 - 12 hours (the

period depends on the type of cover-

slip mountant).