Leica Biosystems CV5030 User Manual

Page 58

Advertising
background image

58

Instructions for Use V 3.0 RevE – 04/2013

5. Operation

Mountant Stroke – stroke length of the coverslip mountant

Mountant Volume - volume of the coverslip mountant

The opening time for the dispenser valve is set in

increments of 1:

Label field

stroke length

The quantity and stroke length of the coverslip

mountant is adapted to the cover slip length.

Fig. 36

Adjustment options for the opening time:
In increments of 1 from 1 (short) to 9 (long).

1 = less coverslip mountant to 9 = maximum quantity of coverslip mountant

The quantity of the coverslip mountant applied also depends on the composition of the mountant

(e.g. viscosity) and the selected dispenser needle (different diameters).

р

р

Mountant Stroke: 40 = 40 mm

Cover slip length

Mountant Stroke: 50 = 50 mm

Cover slip length

Mountant Stroke: 55 = 55 mm

Cover slip length

Mountant Stroke: 60 = 60 mm

Cover slip length

5.10 Menu A – parameter settings (continued)