Instructions for Use V 3.0 RevE – 04/2013

6.

Workstation operation

6.3 Important instructions for operation as workstation

Fig. 44

83

81

The deep bath, which can be identified by side

notches (see Fig.

64

), can be used in workstation

mode in combination with a silver/blue transfer

arm (

Fig. 45

) only. The transfer arm is in the

rear part of the TS5015 / TS5025 Transfer Sta-

tions. Before using the deep bath, please check

whether the silver/blue transfer arm is available.

Older instruments are equipped with a silver/sil-

ver transfer arm. The deep bath may not be used

in combination with this transfer arm!

• Specimen slide holders loaded man-

ually into the CV5030 must always be

removed immediately after process-

ing.

• The unloading chute (

Fig. 44

, 81) of

the transfer stations should be emp-

tied after the "TS FULL" message ap-

pears. At the latest, this should be

done after a signal tone that occurs

after the third specimen slide holder

and the "PAUSED" and "TS FULL"

messages.

• Ensure that the coverslip magazine

in the CV5030 is filled regularly.

This will prevent complete empty-

ing of the coverslip magazine, which

would cause the robotic coverslip-

per to stop working and go into

pause mode.

• Full output magazines must be re-

moved regularly, as otherwise con-

tinuous workstation operation is not

guaranteed.

Fig. 45