Leica CV5030 – Robotic Coverslipper

4. Instrument Setup

4.10 Installing the accessories

(continued)

62

61

Fig. 25

Fig. 27

63

64

Fig. 26

65

66

67

68

Coverslip magazine
• Select and fill a cover slip magazine (62)

for the desired cover slip type (22 or 24 mm

width).

• Tap the cover slip magazine on the edge that

points toward the front of the instrument af-

ter insertion (this ensures that all of the cover

slips are positioned correctly). Then insert the

cover slip magazine into the cover slip maga-

zine holder by this edge and then allow the

magazine to lock in place on the leaf spring

(61) (

Fig. 25

).

Output magazine
• Fill the output station (64) with the desired

number of output magazines (63) (

Fig. 26

). Up

to two size 30 output magazines or three size

20 output magazines fit in the output station.

• Operating the instrument requires that at least

one output magazine is inserted.

Loading bath
• Open the loading door (65) and pull out the

loading drawer (66).

• Insert the loading bath (68) into the drawer and

insert the provided bath insert (

Fig. 27,

67).

• The standard scope of delivery includes the

bath insert for the Leica 30 specimen slide

holder. The respective associated insert must

be used when using other specimen slide

holders (Leica metal 20 specimen slide holder,

Leica Sakura-type plastic 20 specimen slide

holder or specimen slide holders from other

manufacturers (

refer to the Optional Acces-

sories chapter

).