59

Leica CV5030 – Robotic Coverslipper

5. Operation

Stroke Length Corr - stroke length correction of the coverslip mountant

Mountant type - dispensing pressure

An increase of "1" increases the pressure by 100 mbar.

The pressure has to be adapted based on the coverslip mountant used (different viscosi-

ties) and the dispenser needle used (different diameters).

The stroke length for the coverslip mountant set by

the STK (length of the coverslip mountant stroke)

parameter can be shortened by changing LEN (90)

(length correction for the coverslip mountant stroke).

The reduction only applies from the label field area to

the middle of the specimen slide.
The factory preset value is 0.

The pressure needed to dispense the coverslip moun-

tant is set in increments of 1:
1 = 100 mbar/lowest pressure to

10 = 1000 mbar/highest pressure

р

р

In increments of 10 from:

0 = no change

to

-100 = maximum change

Fig. 37

90

The application

quantity is not

changed, but only the

length of the cover-

slip mountant stroke.

5.10 Menu A – parameter settings (continued)

Label field

STROKE LENGTH CORRECTION
(length correction for the coverslip mountant application)
MOUNTANT STROKE
(length of the coverslip mountant application)
STROKE POSITION CORRECTION
(starting point of the coverslip mountant application)
COVERSLIP POSITION CORRECTION
(cover slip placement position)