Leica CV5030 – Robotic Coverslipper

4. Instrument Setup

• After the dispenser needle cleaner is installed,

it must be aligned relative to the dispenser

needle (13).

• To do so, unscrew the screw (7) using the

No. 3 Allen key (6) (

Fig. 20/1

) and align the dis-

penser needle cleaner by moving the bracket

(3) sideways until the dispenser needles (13)

move into the correct position in the middle of

the cleaning port (

Fig. 21

).

• Finally, ensure that the gripper (14) does not

collide with the dispenser needle cleaner

(

Fig. 22

) during its movement or that of the

transported specimen slide (18).

• If the edge of the specimen slide touches the

lid of the brush holder, you can carry out a fine

adjustment after unscrewing the screws (9) on

the housing wall (

Fig. 19/2

). To do so, use the

available play in the bores.

13

14

18

Alignment to the dispenser needle

Fig. 22

Fig. 21

Checking for freedom of movement

7

3

Fill the container with 5 ml of solvent

using the provided plastic pipette

(

Fig. 17,

5). Regularly check the level of

the container. The solvent being used

must be compatible with the coverslip

mountant.