49

Leica CV5030 – Robotic Coverslipper

5.6 Interrupting the coverslipping operation (continued)

5. Operation

If the Pick & Place module loses or breaks a cover slip on the way to the

specimen slide that needs to be covered, the instrument automatically

switches to PAUSE mode. In PAUSE mode, the Pick & Place module can be

moved forward as necessary, toward the user, to make the specimen slide

accessible.
• The READY LED is illuminated in red, the PAUSE LED is green. The dis-

play shows the message "PAUSED CS DROPPED", accompanied by a

signal tone that shows that coverslipping cannot continue.

• Confirm the signal tone with RESPOND.
• Remove the cover slip magazine carefully.

• Check the suction cups of the Pick & Place module for adhesive residue

and, if necessary, clean them using a lint-free cloth moistened with

compatible solvent or replace them with new ones (

see Chap. 7.5.9

).

• Re-insert the cover slip magazine.
• Move the dispenser back into working position.
• Press the START key to resume the coverslipping operation.

• Carefully pull the Pick & Place module forwards (

see mark in Fig. 31

) and

remove the specimen slide to be to be coverslipped from the specimen

slide outfeed.

• Carefully remove the lost cover slip from the instrument.

Fig. 31

Cover slip missing

Put the dispenser into park position to prevent damage to the spec-

imen and/or dispenser needle.

Coverslip the removed specimen slide manually or, if the coverslip

mountant has already begun to dry, immerse the specimen slide in

a compatible solvent, remove the coverslip mountant and re-insert

the specimen slide into the instrument for coverslipping.