Operation – Leica Biosystems CV5030 User Manual

Page 69

Advertising
background image

69

Leica CV5030 – Robotic Coverslipper

5. Operation

5.13.1 Procedure (continued):

Important note:

If TYP is at the maximum setting (10 - highest pressure), it is possible to generate small air

bubbles in the dispenser system. This depends on specific properties of the respective cover-

slip mountant and the selection of dispenser needle (frequently occurring with small-diameter

needles). Therefore it is advisable to begin with the mid-range parameter settings when deter-

mining the optimum parameter sets.

Unsuitable procedures:

Dispenser needle No. 21

TYP: 10

VOL: 1 or 2 incremental increase.

Suitable procedures:

Dispenser needle No. 21

TYP: 4 or 5 incremental increase/reduction, with parallel incremental increase/reduction of VOL

VOL: 4 or 5 incremental increase/reduction, with parallel incremental increase/reduction of TYP

If a suitable coverslip mountant application quantity cannot be attained, switch to a dispenser

needle with a different diameter.