6.

Workstation operation

6.4 Interrupting workstation operation

• The PAUSE and STOP buttons have mostly the

same functions outlined in

Chapter 5.6

.

• If the coverslipping operation is interrupted

using the STOP button, the coverslipper must

be initialized. To do so, press the START button.

• After the coverslipping operation is interrupted

using the STOP button, the specimen slide

holder, which is still in the instrument, must

be removed from the loading bath manually

after processing is finished. In this case, it

is not transported automatically back to the

transfer station.

• In case of power failure or if the instruments are switched off, specimen slide holders can

remain in different positions. Therefore, if this error occurs, the following items must be

checked:

- Transfer station with all transfer facilities

- Loading bath of the coverslipper

• These specimen slide holders must be removed manually to prevent collisions after resuming

workstation operation.

• The specimen slide holders that are left behind must be placed in the coverslipper manually

for further processing and removed after processing.

So that you do not disturb workstation operation, you should process specimen slide

holders that come from the slide stainer first. Only then should you process the specimen

slide holders to be loaded manually in the coverslipper.

If the instrument is operated as a workstation, the user must ensure that before reinitial-

ization, no rack is in the transfer station.