1 removing or installing the transport anchors, Instrument setup – Leica Biosystems CV5030 User Manual

Page 21

Advertising
background image

21

Leica CV5030 – Robotic Coverslipper

4. Instrument Setup

28

4.3.1 Removing or installing the transport anchors

• Open the service door (28) on the left

side of the instrument and remove the

foam (50).

24

25

23

22

Fig. 5

25

Fig. 4

50

• To transport the instrument, reinstall the transport anchors in the opposite order.

1. Remove the transport anchor (24) for the Pick & Place module.
2. Remove the transport anchor (25) for the gripper. The gripper slowly moves downward in the

process.

• Using the provided No. 3 Allen key, unscrew the screws (22) and (23) for the two red transport

anchors (24) and (25) (

see Fig. 5

):

Transport anchor (24) for

the Pick & Place module

Transport anchor for the

gripper (25)

24

23

We recommend keeping the transport anchors and screws close to the instrument in the

provided clear plastic bag.

Advertising