Leica CV5030 – Robotic Coverslipper

8. Malfunctions and Troubleshooting

8.1 Error codes

Display

Cause

Troubleshooting

Error 301

SLIDER BLOCKED

The specimen slide outfeed

is blocked.

Check the specimen slide outfeed (

refer to Chapter 7.5.11

) and

output magazine (

refer to Chapter 7.6.3

). Remove the cause of the

blockage (glued-on material due to coverslip mountant residue) if

necessary (

refer to Chapter 7.6.3

). Switch the instrument off and

back on and proceed as described in

Chapter 5.3

.

Error 305

GR-X BLOCKED

The horizontal movement

of the gripper tongs is

blocked.

Remove the cause of the blockage. If the gripper tongs have

gripped a specimen slide, use RELEASE SLIDE to open the gripper

tongs and manually remove the specimen slide. Then switch the

instrument off and on and proceed as described in

Chapter 5.3

.

Ensure that any adhesive residue has been cleaned from the

gripper jaws (adhesive from leftover labels sensitive to solvents)

(

refer to Chapter 7.6.3

).

Error 306

GR-Z BLOCKED

The gripper is blocked in its

vertical movement.

Remove the cause of the blockage. If the gripper tongs have

gripped a specimen slide, use RELEASE SLIDE to open the gripper

tongs and manually remove the specimen slide. Then switch the

instrument off and on and proceed as described in

Chapter 5.3

.

Error 312

TS-X BLOCKED

The transfer arm of the

TS5025/5015 Transfer Sta-

tion is blocked in its left/

right movement (X-axis).

Remove the cause of the blockage (specimen slide holder) and

inspect the transfer arm of the TS5025/TS5015. Remove empty

specimen slide holders from the discharge chute of the transfer

station. Switch the instrument off and on. Proceed further as de-

scribed in

Chapter 5.3

. Place any removed specimen slide holders

into the coverslipper and have them coverslipped (thus outside of

the workstation operation).

All of the error messages with the corresponding actions for troubleshooting are listed below.

If faults occur that cannot be corrected with the procedures recommended in the table, or if

they occur again, contact the responsible Leica Service contact person.

In case of errors, generally the user's first step is to secure the specimens located at different

positions in the instrument/workstation while ensuring personal safety.