Instructions for Use V 3.0 RevE – 04/2013

If the number of cover slips falls below the limit, the message "COVERSLIPS

LOW" appears on the display.

The instrument continues to operate until there are no more cover slips

in the cover slip magazine. In this case, the messages "PAUSED" and "CS

EMPTY" are shown on the display and a signal tone is emitted.

• Switch off the signal tone with RESPOND.

• Fill the cover slip magazine and then continue the coverslipping operation

by pressing START.

Empty coverslip magazine

Cover slips damaged

• The cover slip sensor between the suction cups in the Pick & Place

module automatically detects defective cover slips.

• The coverslipping operation continues automatically if an undamaged

cover slip is picked up next.

• The coverslipping operation is interrupted if three damaged cover slips

in a row are detected and discarded. The messages "PAUSED" and

"COVERSLIP BROKEN" are displayed accompanied by a signal tone.

• Confirm the signal tone with RESPOND.
• Check the cover slips in the cover slip magazine for further damage and

fill with new cover slips if necessary.

• Then, continue the coverslipping operation by pressing START.

5.6 Interrupting the coverslipping operation (continued)