Instrument setup – Leica Biosystems CV5030 User Manual
Page 30
30
Instructions for Use V 3.0 RevE – 04/2013
4.
Instrument Setup
11
9
4
12
①
Assembly
To install the holder for the
dispenser needle cleaner, you
must first remove the transport
anchors for the Pick & Place
module (see
Chap. 4.3.1, Fig. 5
).
For installation, use the thread-
ed holes (11) of the transport
anchor (
see Fig. 19/1
).
• First, fasten the mounting bracket (4) in the
bores (11) in the housing wall (12) using the
two screws (9) (
Fig. 19/2
). Make sure that the
mounting bracket is parallel to the edge of the
housing wall (
ellipse, in Fig. 19/2
).
• Then, fasten the holder (3) on the mounting
bracket using the screw (7) and shim (8).
• Insert the completely assembled dispenser
needle cleaner (1) into the holder as shown
in
Fig. 20/2
. Press down to ensure that the two
lateral mounting clips (17) audibly engage in
the notches of the bracket.
3
7, 8
➁
1
6
➁
Fig. 19
①
17
Fig. 20
4.8 Dispenser needle cleaner (nozzle cleaner) (continued)