Operation – Leica Biosystems CV5030 User Manual

Page 45

Advertising
background image

45

Leica CV5030 – Robotic Coverslipper

• In an emergency or if the coverslipping operation is to be canceled

quickly, press the STOP button.

Then, the coverslipping operation is interrupted immediately.

• The message "STOPPED" is displayed accompanied by a signal tone.

• Confirm the signal tone with RESPOND.

Emergency stop using STOP.

5. Operation

• Using the RELEASE COVER SLIP and RELEASE SLIDE buttons, you can

release any cover slip that has been picked up or specimen slide that

has been gripped.

• Only press the START button to continue working after doing so. The

instrument then initializes.

• Continue as outlined in

Chapter 5.3

.

5.6 Interrupting the coverslipping operation (continued)

ò

Remove all glass parts or other objects away from the CV5030

working area after pressing STOP. Only then may START be

actuated.

Proceed carefully when removing a specimen slide that has been

gripped. Do not manually open the grippers under any circum-

stances.

Immediately remove all glass parts, broken glass (e.g. cover slips)

or other objects from the instrument's working area.

The output magazines are moved to the output position automati-

cally after pressing START. All remaining specimen slides in the

output magazines absolutely must be removed.