Operation – Leica Biosystems CV5030 User Manual

Page 48

Advertising
background image

48

Instructions for Use V 3.0 RevE – 04/2013

If the dispenser system cannot reach the intended pressure or vacuum

after a certain time, the instrument automatically switches to PAUSE mode.
• The READY LED is illuminated in red, the PAUSE LED is green. Depend-

ing on the error that has occurred, the display reads "PAUSED BOTTLE

VACUUM" or "PAUSED BOTTLE PRESSURE", accompanied by an signal

tone indicating that coverslipping cannot continue.

• Confirm and switch off the signal tone with RESPOND.

• Check the entire dispenser system for leaks.

To do so, check whether:
- The O-ring is inserted in the cover of the dispenser group and intact.

- The connections of the air hose are tight.

- Hardened adhesive residue is on the dispenser needle.
• If the dispenser is still in the working position, the message "PRIME

POSITION" appears on the display. Put the dispenser into the rest posi-

tion and continue as outlined in

Chapter 5.3

.

5.6 Interrupting the coverslipping operation (continued)

5. Operation

Pressure or vacuum cannot be reached