• Open the hood and place a suitable spirit level

(26) onto the work surface as shown in

Fig. 6

.

• The instrument can be leveled in both direc-

tions by screwing or unscrewing the instru-

ment feet (27) (

detail 2 in Fig. 6

).

4.4 Leveling the instrument

Fig. 6

27

• Move the instrument to its final location on the

work surface.

When doing so, ensure that all four feet are

on the work surface.

Detail 2

4.

Instrument Setup

4.5 Exhaust system

26

We recommend operating the robotic coverslipper with an exhaust hose and connection to ex-

ternal laboratory extraction or under a suitable fume hood. The associated active carbon filter

is intended to be used in a supporting role for this. We recommend replacing the active carbon

filter every three months. Please note that the active carbon filter, when used alone, can filter

only a limited number of harmful vapors (e.g. xylene). The setup location can differ substan-

tially in terms of solvent loading, room ventilation, room/ambient temperature, room size, etc.

In case of doubt, the laboratory owner/operator must have measurements conducted on site to

ensure that the legal limits for solvent vapors are not being exceeded.