Instrument setup – Leica Biosystems CV5030 User Manual
Page 25
25
Leica CV5030 – Robotic Coverslipper
4. Instrument Setup
• Select the dispenser needle to be used for cov-
erslipping from the scope of delivery (
Fig. 12
).
• Insert the dispenser needle (46) into the dis-
penser needle holder (48) from below and turn
the dispenser needle by 45° until one corner
(47 in
Fig. 12
) is over the retaining plate (49)
(
Fig. 11
).
44
Inserting the dispenser needle
(
Fig. 11
)
Fig. 11
4.6 Installing the dispenser group
(continued)
Fig. 10
Fig. 12
47
Dispenser needles
• Finally, insert the cable and the air hose into
the holder (45) provided (
Fig. 10
)
45
48
46
49
21 G
16 G
18 G
20 G
Ensure correct positioning for the dis-
penser needle, since the dispenser
needle can loosen during the cover-
slipping operation. Leaks can create
bubbles during coverslipping.