Leica CV5030 – Robotic Coverslipper

4. Instrument Setup

• Select the dispenser needle to be used for cov-

erslipping from the scope of delivery (

Fig. 12

).

• Insert the dispenser needle (46) into the dis-

penser needle holder (48) from below and turn

the dispenser needle by 45° until one corner

(47 in

Fig. 12

) is over the retaining plate (49)

(

Fig. 11

).

44

Inserting the dispenser needle

(

Fig. 11

)

Fig. 11

4.6 Installing the dispenser group

(continued)

Fig. 10

Fig. 12

47

Dispenser needles

• Finally, insert the cable and the air hose into

the holder (45) provided (

Fig. 10

)

45

48

46

49

21 G

16 G

18 G

20 G

Ensure correct positioning for the dis-

penser needle, since the dispenser

needle can loosen during the cover-

slipping operation. Leaks can create

bubbles during coverslipping.