Workstation operation – Leica Biosystems CV5030 User Manual

Page 77

Advertising
background image

77

Leica CV5030 – Robotic Coverslipper

6. Workstation operation

6.3 Important instructions for operation as workstation (continued):

Important instructions for operation of different specimen slide holders

• Apart from workstation operation, the robotic

coverslipper can be manually loaded with

specimen slide holders from other manufac-

turers, or with other Leica specimen slide

holders, with use of a compatible bath insert

only.

• After coverslipping is concluded, ensure that

the specimen slide holder is removed once

again. Likewise, the replaced bath insert is to

be removed before starting a new workstation

operation.

The default specimen slide hold-

ers intended for workstation op-

eration are Leica plastic holders

for 30 specimen slides (Order

No. 14 0475 33643).

Use of other Leica specimen slide

holders (e.g. Leica plastic holders

for 20 specimen slides) or speci-

men slide holders from other man-

ufacturers is not permissible in

workstation operation.

Specimen slide holders from other

manufacturers are only permis-

sible when they have been tested

on the instrument. These specimen

slide holders, or compatible bath

inserts, are listed in

Chapter 9 -

"Optional Accessories"

. Impermis-

sible specimen slide holders may

in some cases not be recognized

by the instrument and a collision

may result.