5.6 Interrupting the coverslipping operation

Press PAUSE to interrupt the coverslipping operation.

• The currently processed specimen slide is completely coverslipped

and pushed into the output magazine, accompanied by the message

"PAUSING".

• If this step is ended, the message "PAUSED" is shown in the display.

A signal tone is emitted at the same time.

• Confirm the signal tone by pressing the RESPOND button.

• Use the START button to continue the coverslipping operation.

ò

5. Operation

Whenever the instrument is to pause only briefly (e.g. to refill the

cover slip magazine), use PAUSE to interrupt the instrument with-

out having to run through a complete initialization and prime cycle.

If a just partly filled output magazine is to be removed, press LIFT

and proceed as outlined under Chap. 5.5.

In PAUSE mode, pressing and holding the RESPOND button

(>2 sec) allows you to change coverslipping parameters or select

another parameter set (SET 1 through 4) (

see Chap. 5.8

).