Leica CV5030 – Robotic Coverslipper

4. Instrument Setup

4.7.2. Setting the needle height

Fig. 15

39

35

The needle height has to be corrected after rein-

serting the dispenser needle.

Setting the needle height:

• Reset the dispenser (35) from the working

position (

see Fig. 16,

57) to the rest position

(

see Fig. 15,

39).

• There is a screw (

see Fig. 16

, 54) in the work-

ing position. This determines the distance

between the dispenser needle and specimen

slide.

• The height of the dispenser can be changed

by rotating the screw using a No. 3 Allen key

(56) (standard delivery):

-

Rotating clockwise decreases the dis-

tance.

-

Rotating counterclockwise increases the

distance.

• Keep rotating the screw clockwise until the

dispenser needle is touching the highest point

of the specimen slide outfeed (

see Fig. 14,

55)

(distance = 0 mm). This can be checked by

putting the dispenser into the working position.