Leica CV5030 – Robotic Coverslipper
4. Instrument Setup
4.7.2. Setting the needle height
Fig. 15
39
35
The needle height has to be corrected after rein-
serting the dispenser needle.
Setting the needle height:
• Reset the dispenser (35) from the working
position (
see Fig. 16,
57) to the rest position
(
see Fig. 15,
39).
• There is a screw (
see Fig. 16
, 54) in the work-
ing position. This determines the distance
between the dispenser needle and specimen
slide.
• The height of the dispenser can be changed
by rotating the screw using a No. 3 Allen key
(56) (standard delivery):
-
Rotating clockwise decreases the dis-
tance.
-
Rotating counterclockwise increases the
distance.
• Keep rotating the screw clockwise until the
dispenser needle is touching the highest point
of the specimen slide outfeed (
see Fig. 14,
55)
(distance = 0 mm). This can be checked by
putting the dispenser into the working position.
Fig. 16
57
56