Instructions for Use V 3.0 RevE – 04/2013

5. Operation

• Here, the following parameters must be matched to each other:

STP (Stroke position - starting point for the coverslip mountant stroke)

CSP (Cover slip placement position) and

LEN (Stroke length correction for the coverslip mountant stroke)

STP and CSP must be matched with each other, since otherwise there is the danger that

the coverslip mountant could drip over the specimen slide edge into the instrument at the

stroke start point, or become stuck to the suction cup due to poor coverslip placement. In

both cases, continuing smooth instrument operation cannot be ensured.

8. Fine-tuning the adhesive and coverslip support positioning:

Fig. 40

5.13.1 Procedure (continued):

Label field

STROKE LENGTH CORRECTION
(length correction for the coverslip mountant application)
MOUNTANT STROKE
(length of the coverslip mountant application)
STROKE POSITION CORRECTION
(starting point of the coverslip mountant application)
COVERSLIP POSITION CORRECTION
(cover slip placement position)