Leica CV5030 – Robotic Coverslipper

74

5.4 Brief inspection before starting the

coverslipping operation

65

68

Fig. 30

The following points have to be checked again

before starting the coverslipping operation:

• The level of the loading bath must be checked

before starting the coverslipping process. Fill

the loading bath with the corresponding sol-

vent. The level should reach up to the label

field of the specimen slide when a specimen

slide holder (74) is inserted.

• When inserting the loaded specimen slide

holder, ensure that the specimen slide re-

ceiving the cover slip has the side with the

specimen pointing towards the user (instru-

ment front).

5. Operation

• Push the loading drawer (68) with the loading

bath back into the instrument; close the load-

ing door (65) afterwards.

• Furthermore, check and, if necessary, correct

the following before starting the coverslipping

operation:

- Filling of the dispenser needle cleaner.
- Level of the coverslip mountant bottle.
- Sufficient filling and correct positioning of the

cover slip magazine.

- Presence of the output magazines.
- Selection of the desired parameter set (

refer to

Chapter 5.9

).