Instructions for Use V 3.0 RevE – 04/2013

5.2 Key functions for instrument operation (continued)

5. Operation

RELEASE COVER SLIP

RELEASE COVER SLIP ventilates the Pick & Place module's suction

cups.

The cover slip is released.

The function is disabled while the BUSY LED is illuminated in green.

RELEASE SLIDE

Pressing RELEASE SLIDE opens the gripper jaws as long as the

button is held down.

The function is disabled while the BUSY LED is illuminated in green.

PRIME

Pressing the PRIME key opens the dispensing valve as long as the

key is pressed.

The function is disabled while the BUSY LED is illuminated in green.

VENT MOUNTANT

VENT MOUNTANT generates a vacuum in the coverslip mountant

bottle, thereby assisting the coverslip mountant with degassing.

This function is disabled if the BUSY LED is illuminated in green.

Exit

This button is used to exit the configuration menu.