Operation – Leica Biosystems CV5030 User Manual

Page 54

Advertising
background image

54

Instructions for Use V 3.0 RevE – 04/2013

5.7 Display indicators and instructions (continued)

5. Operation

Display

Cause

Action

The Pick & Place module loses

the coverslip on the way to the

coverslipping position.

As soon as the display reads "CS

DROPPED", the instrument goes

automatically into PAUSE mode.

Notice: Move the dispenser into

the park position, move the Pick

& Place module towards the

user and remove the specimen

slide. You can now coverslip

the specimen slide manually.

As an alternative, immerse the

specimen slide in a compatible

solvent and remove the cover-

slip mountant, then re-insert the

specimen slide into the instru-

ment for coverslipping.

1. The suction cups, the cover

slip sensor and/or the skids on

the Pick & Place module are

stuck in place with coverslip

mountant.

1. Clean the suction cups, cover

slip sensor and/or the skids of

the Pick & Place module using

a lint-free cloth moistened

with compatible solvent or

replace them with new ones

(

see Chap. 7.5.9

).

2. The suction cups are de-

formed.

2. Replace the suction cups with

new ones (

see Chap. 7.5.9

).

3. The cover slip magazine is

inserted incorrectly.

3. Check that the cover slip

magazine is seated correctly.

4. The suction cups are not

correctly seated on the Pick &

Place module.

4. Check that the suction cups

are seated correctly; if neces-

sary, pull them off and reat-

tach.