Instrument setup – Leica Biosystems CV5030 User Manual

Page 23

Advertising
background image

23

Leica CV5030 – Robotic Coverslipper

4. Instrument Setup

Inserting the active carbon filter
•
Unscrew the slotted screw (34) at the service

door (28) on the left side of the instrument and

open the service door to the left (

Fig. 7

).

• Unscrew the screw (29) of the filter cover (30)

using a No. 3 Allen key and pivot the cover

upwards.

• The insertion date can be written down on the

adhesive label (31).

• Insert the filter (32), close the cover (30) and

secure it back in place using the screw (29).

• Finally, close the service door and retighten

the slotted screw.

Attaching the exhaust hose (detail in

Fig. 7

)

• For installation, push the exhaust hose (70)

onto the pipe (33) as far as it will go (

refer to

the yellow arrow in the detailed figure

); the

pipe is on the rear side of the service door (28).

4.5 Exhaust system

(continued)

Fig. 7

28

29

31

70

Filter label

32

31

34

33

Detail:

30

Note:

Fig. 7

detail: Assembly of the exhaust

hose (70). The filter cover (30) and filter

(32) have been removed to show the

connecting piece (33).