Instructions for Use V 3.0 RevE – 04/2013

4.

Instrument Setup

4.7 Aligning the dispenser needle height relative to the specimen slide outfeed

Fig. 13

The height of the dispenser needle must be

aligned to the specimen slide correctly so that

no air bubbles are generated while the mountant

is applied. The dispenser needle height must not

be set too low in order to avoid damage to the

specimen on the specimen slide.
Aligning the dispenser needle:
• Switch off the instrument and unplug it.

Open the service door (28) as outlined in

Chap. 4.6 (Fig. 7)

.

• Remove the cover slip catch tray that may

already be attached at the specimen slide

outfeed.

• Move the dispenser with the dispenser needle

into the working position.

• There is a pulley (51) that moves the specimen

slide outfeed (52) in the area near the opened

service door. This can be used to move the

specimen slide outfeed to the left and right

(

see yellow arrow, Fig. 13,

51).

• The dispenser needle can be moved back-

ward and forward in the working position

(

see Fig. 13,

53) carefully by hand using the

Pick & Place module.

55

Detail:

51

52

28

4.7.1 Checking needle height

53

Fig. 14

55

46

• Now align the highest point of the specimen

slide outfeed (see

Fig. 14,

55) and the tip of the

dispenser needle so that they are touching.

Caution!

Do not cause the Pick & Place module

to make any downward motion.