background image

28

Instructions for Use V 3.0 RevE – 04/2013

4

Instrument Setup

4.7.2 Setting the needle height

(continued)

• The dispenser is reset into the rest position if

the 0 mm distance is set correctly.

• Now carry out a three-quarters counterclock-

wise revolution using the No. 3 Allen key.

• This achieves the optimum distance of 0.75 -

0.8 mm.

• The set height of the dispenser needle can be

inspected again by a visual inspection from the

front (

see Fig. 14

).

• Then close the service door, screw it down

tight and reconnect the instrument to the

power supply.