4instrument setup – Leica Biosystems CV5030 User Manual
Page 28
Advertising
28
Instructions for Use V 3.0 RevE – 04/2013
4
Instrument Setup
4.7.2 Setting the needle height
(continued)
• The dispenser is reset into the rest position if
the 0 mm distance is set correctly.
• Now carry out a three-quarters counterclock-
wise revolution using the No. 3 Allen key.
• This achieves the optimum distance of 0.75 -
0.8 mm.
• The set height of the dispenser needle can be
inspected again by a visual inspection from the
front (
see Fig. 14
).
• Then close the service door, screw it down
tight and reconnect the instrument to the
power supply.