8 procedure for changing the cover slip mountant, Procedure for changing the cover slip mountant, Cleaning and maintenance – Leica Biosystems CV5030 User Manual

Page 88

Advertising
background image

88

Instructions for Use V 3.0 RevE – 04/2013

7.

Cleaning and Maintenance

7.8 Procedure for changing the cover slip mountant

7.8.1 Changing from one xylene-based

coverslip mountant to another

• Remove the mountant bottle.

• Insert a second glass bottle with 150 ml of

xylene and rinse/prime it (see

Chap. 7.3

B).

• Insert a mountant bottle with the new moun-

tant.

• Run a prime cycle. Hold down the PRIME

button until the hose is completely free of air

bubbles and filled with coverslip mountant.

7.8.2 Changing from a xylene substitute

coverslip mountant to a xylene-based

mountant

• Remove the mountant bottle.

• Insert a second glass bottle with 150 ml of

xylene substitute and rinse/prime it (see

Chap. 7.3

B).

• Empty the second glass bottle and fill it with 150 ml

of 100 % ethanol before inserting it and rinsing.

• Empty the second glass bottle and fill it with

150 ml of xylene before inserting it and rinsing/

priming.

• Insert a mountant bottle with the new mountant.

• Run a prime cycle. Hold down the PRIME

button until the hose is completely free of air

bubbles and filled with coverslip mountant.

7.8.3 Changing from a xylene-based

coverslip mountant to a xylene

substitute

The sequence of individual substanc-

es has to be followed to avoid milky/

cloudy clumping in the hose and the

dispenser group.

All parts that were filled with xylene

must be converted to the substitute

(= loading bath, glass vial, dispenser

needle cleaner)!

• Remove the mountant bottle.

• Insert a second glass bottle with 150 ml of

xylene and rinse/prime it (see

Chap. 7.3

B).

• Empty the second glass bottle and fill it with

150 ml of 100 % ethanol before inserting it and

rinsing.

• Empty the second glass bottle and fill it with

150 ml of xylene substitute before inserting it

and rinsing/priming.

• Insert a mountant bottle with the new, substi-

tute-compatible mountant.

• Run a prime cycle. Hold down the PRIME

button until the hose is completely free of air

bubbles and filled with coverslip mountant.

Before switching the mountant, switch

off the instrument and disconnect it from

the power supply! Remove the loading

bath and specimen slide holders from

the instrument before switching it off.

Due to the high pressure when rinsing

with solvent, it is necessary to wear

gloves, safety goggles and appropri-

ate protective clothing when changing

the cover slip mountant! Select an ad-

equately sized container to catch the

rinsing residue.