Leak-test – Fluke Biomedical VT Plus HF User Manual

Page 100

Advertising
background image

VT Plus HF

Operators Manual

4-16

Leak-Test

The Leak test is used for measuring the leak rate of a sealed vessel or test lung over a

period of time.

Setting Leak Test Parameters

The Leak test setup screen, shown in Figure 4-11, allows the user to choose the desired

pressure channel on which to run the test.

fec093.bmp

Figure 4-11. Leak Test: Setup Screen

Also, the displayed pressure units can be changed. If a compliance value is entered for

the system being tested, the volume lost during the test is calculated. Enter the test time to

set the length of time that the Leak test is to run. At the end of that amount of time, the

Leak test automatically stops.

Running the Leak Test

To run the Leak test:
1. Press the

START

button to begin the Leak test. The screen changes, as shown in

Figure 4-12: