Zeroing, Manual zeroing, Automatic zeroing – Fluke Biomedical VT Plus HF User Manual

Page 114

Advertising
background image

VT Plus HF

Operators Manual

4-30

Zeroing

The Analyzer can be zeroed either automatically or manually. Zeroing corrects the flow

and pressure sensors for offset to get a more accurate reading.
To switch the Analyzer between automatic and manual zeroing:
1. From the

Setup

menu, select the

Zeroing

menu. The current zeroing mode has a

check next to its name.

2. Select the desired zeroing mode—Manual or Automatic.

Manual Zeroing

When the Analyzer is in manual zero mode, no automatic zeros occur. Periodically, it is a

good idea to zero the signals on the Analyzer to null any offset errors in the flow and

pressure sensors.
The Analyzer has two ways to perform a manual zeroing:
1. Select

Zero

from the menu.

2. Choose

Zero All

or select the signal to be zeroed from the given list.

fec114.bmp

Or
1. Click the

Zeroing Speed

button on the toolbar.

2. Choose the preferred plot from the given list.
3. Click the

Zeroing

button; this is the same as selecting

Zero All

.

Automatic Zeroing

The Analyzer can be set to automatically zero flow and airway pressure periodically. This

nulls offset errors in the pressure transducers. During an auto-zero, the Analyzer uses

pneumatic valves to temporally switch flow ports for the zero. Do not use this option if

there is flow in both flow ports. A zero can be forced by manually zeroing a signal, even

if automatic zeroing is turned on.

Advertising