Fuse Replacement

The Analyzer fuses may have to be replaced. The fuse compartment is located on the side

panel of the instrument, as shown in Figure 5-1.

fec122.eps

Figure 5-1. Fuse Compartment

To replace the fuses:
1. Power off Analyzer and unplug the main power cord.
2. Using a small, flat-blade screwdriver or similar tool, carefully pry off the cover to the

fuse compartment. The fuse block assembly is attached to the back of the cover.

3. Pull out the fuse block assembly and carefully lift the old fuses from the fuse slots.
4. Replace the fuses (two 0.5 A Slo-Blo fuses, Fluke part #46026).

XW

Warning

To avoid possible fire, electrical shock, or personal injury, use
only the recommended fuse.

5. Replace the cover/fuse block in the fuse compartment.
6. Plug in the Analyzer.