Errors, Restore defaults – Fluke Biomedical VT Plus HF User Manual

Page 46

Advertising
background image

VT Plus HF

Operators Manual

2-18

Errors

When an error is made, the Analyzer signals it with a beep. This option adjusts the sound

level of that beep.
To adjust the sound level of errors:
1. Highlight

Errors

and press the

MODIFY

soft key to toggle between

Low

,

High

, or

Off

.

2. Press the

ENTER

soft key to save the change or the

ESCAPE

soft key to cancel the

change.

3. Press the

BACK

soft key to return to the

Settings

menu.

Restore Defaults

This option returns all values to the factory defaults listed below, but does not change the

device calibration.
To restore defaults:
From the

System

menu, highlight

Restore Defaults

. The Analyzer restores system

values to the default, displays a Restoring to defaults. Please wait message.
After a few seconds, a beep indicates defaults are restored, and the display returns to the

System

menu. Factory defaults are shown in Table 2-4.

Table 2-4. Factory Defaults

System

Date Format

MM/DD/YYYY

Demo Data

Off

Filtering None

Display Cal Date

Yes

Serial Mode

None

Printer Options

Printer Type

None

Graphics

Text Only

Audio Levels

General Low

Keypad Low

Errors

Low

Settings

Gas Settings

Gas Type

Air

Gas Temperature

22

°

C

Ambient Temperature

22

°

C