Connection and Setup

Setting Testing Parameters

2

2-9

1. From the

Gas Settings

menu, highlight

Gas Type

.

2. Press the

MODIFY

soft key until the desired value displays.

3. Press the

BACK

soft key to return to the

Settings

menu.

Gas Temperature

This setting is the temperature of the gas flowing through the high- or low-flow port. The

Analyzer uses this information to improve the accuracy of the flow measurements. All

temperatures are in °C.
To enter a gas temperature:
1. From the

Gas Settings

menu, highlight

Gas Temperature

.

2. Press the

MODIFY

soft key to display the value field.

3. Input the desired value from the keypad.

Note

The Analyzer accepts temperatures between 0 and 99 °C. No decimal

values or fractions are allowed.

4. Press the

ENTER

soft key to save the change or the

ESCAPE

soft key to cancel the

change.

5. Press the

BACK

soft key to return to the

Settings

menu.

Ambient Temperature

This setting is the room temperature air flowing through the high-or low-flow ports.
To change ambient temperature:
1. From the

Gas Settings

menu, highlight

Ambient Temp

.

2. Press the

MODIFY

soft key to display the value field.

3. Input the desired value from the keypad.

Note

The Analyzer accepts temperatures between 0 and 99 °C. No decimal

values or fractions are allowed.

4. Press the

ENTER

soft key to save the change or the

ESCAPE

soft key to cancel the

change.

5. Press the

BACK

soft key to return to the

Settings

menu.

Relative Humidity

This setting is the relative humidity in the air flowing through the high- and low-flow

ports and is used only for STPD and BTPS corrections. The relative humidity is not used

in calculating the effective viscosity on the Analyzer; however, the gas calculator in the

VT Plus PC software does use relative humidity in determining the effective viscosity of

user-defined gases.
To change relative humidity:
1. From the

Gas Settings

menu, highlight

Rel. Humidity

.

2. Press the

MODIFY

soft key to display the value field.