VT Plus HF

Operators Manual

3-22

1. Use the S and T soft keys to highlight

Test Parameter

.

2. Press the

MODIFY

soft key until the desired breath parameter displays. The most

recent measurement of the selected parameter is displayed as the

Current Value

.

3. Use the S and T soft keys to highlight

% Limit

.

4. Press the

MODIFY

soft key, then use the keypad to enter the percent change

representing a deviation from acceptable limits. Percent values range from 1 to 99.

Remember to enter leading zeros for values less than 10. For example, 5 % is entered

as 05 %.

5. Press the

ENTER

soft key.

6. Use the S and T soft keys to highlight

Use Printer

.

7. Press the

MODIFY

soft key. If

Yes

is selected, a report is printed that list all breath

parameters with a time/date stamp every time an incident occurs.

To run the test:
1. Ensure that the ventilator is on and that steady-state conditions are reached. Monitor

the

Current Value

displayed to ensure that the breath parameter is as expected.

Press the

START

soft key

to begin testing. While the test is in progress, the Trend

Test screen displays the parameter being tested, the current and starting values, and

the % difference between the two values. The allowable % deviation limit, the

number of tests taken, and the number of breaths (incidents) in which the limit has

been exceeded is also displayed.

fec059.bmp

2. To terminate the test, press the

STOP

soft key. A sample Trend Test report is shown

below.