Low Pressure (+ and -)

The low-pressure ports connect using standard luer fittings. Note that only the + side can

be used to measure fluid pressure.

W

Caution

To avoid damage to the Analyzer or adverse affects on its
performance, follow these guidelines:

Pressures applied to either side of the low-pressure port
should not exceed ±500 mmHg.

Negative (-) side of the low-pressure port cannot be used to
measure fluid pressure.

Test System Setup

High- or low-flow ranges can be used for ventilator testing. One test setup is measuring

ventilator parameters using a test lung. Either a bi-directional flow mode, as shown in

Figure 2-1 or a unidirectional flow mode, as shown in Figures 2-2 and 2-3, below can be

chosen; however, the bi-directional mode is preferred.

Bi-directional Flow Mode

1. Connect the ventilator, using a Y-piece adapter, to the flow inlet port on the right of

the Analyzer.

2. Connect the flow exhaust port on the left to the test lung using standard breathing

hoses.

O

u

tlet

Inlet

V

entilator

Inspiratory Hose

Expiratory Hose

Analyzer

TEST
LU

N

G

fec002.eps

Figure 2-1. Connecting the Analyzer in a Bi-Directional Flow Mode

3. Ensure that the breath detection mode is set to bi-directional.

a. Go into

Setup

.

b. Choose

Settings

c. Choose

Breath Detect

and modify as needed