Connection and Setup

Connecting the Analyzer

2

2-5

Unidirectional Flow Mode

Alternatively, the Analyzer can be used to measure only inspiratory or expiratory gas

flow, as described below.

Inspiratory Testing

1. Connect the inspiratory hose to the flow port on the right of the Analyzer.
2. Connect the flow exhaust port on the left to the test lung using standard breathing

hoses.

O

u

tlet

Inlet

V

entilator

Inspiratory Hose

Expiratory Hose

TEST

LU

N

G

V

T PLUS HF

fec008.eps

Figure 2-2. Connecting the Analyzer to the Inspiratory Limb of the Breathing Circuit

3. Ensure that the breath detection mode is set to Unidirectional Inspiratory.

a. Go into

Setup

b. Choose

Settings

c. Choose

Breath Detect

and modify as needed

Expiratory Testing

1. Connect the expiratory hose to the flow port on the right side of the Analyzer.
2. Connect the flow exhaust port on the left to the ventilator using standard breathing

hoses.

Inlet

O

u

tlet

V

entilator

Inspiratory Hose

Expiratory Hose

TEST

LU

N

G

V

T PLUS HF

fec009.eps

Figure 2-3. Connecting the Analyzer to the Expiratory Limb of the Breathing Circuit