Utilities, Oxygen sensor calibration – Fluke Biomedical VT Plus HF User Manual

Page 48

Advertising
background image

VT Plus HF

Operators Manual

2-20

Utilities

Note

These settings are required by service technicians in servicing the Analyzer.

To access the utility functions:
1. From the

Setup

screen, highlight

Utilities

and press the

ENTER

soft key. The

Utilities

menu displays:

fec035.bmp

Oxygen Sensor Calibration

Recalibrate the oxygen sensor with 100 % oxygen and dry air at the beginning of any day

in which oxygen concentrations are to be measured.

To calibrate the oxygen sensor:

1. Apply 10 lpm of dry air to the high-flow inlet or exhaust.
2. Use the S and T soft keys to highlight

Oxygen Calibration

.

3. Press the

ENTER

soft key. The

O2 Calibration

screen displays.

fec075.bmp