Fluke Biomedical VT Plus HF User Manual
Page 73

Operation
Special Functions
3
3-23
TREND TEST REPORT
6/13/2000
10:45
VT PLUS HF S/N: 4294967295
Control #: __________________
S/N: __________________
Tech: __________________
Incident Report
Test Parameter:
Minute Vol
Elapsed Time:
0:00:05
Current Value:
40.333 L
Starting Value:
28.686 L
% Difference:
41 %
% Limit:
5 %
# of Tests:
1
# of Incidents:
1
Parameters:
Minute Vol:
40.333 L
Breath Rate:
29 bpm
Tidal Vol:
1384.8 ml
Ex Time:
0.9 sec
PIP:
109.42 LPM
In Time:
1.1 sec
PEP:
113.91 LPM
Ex Hold:
0.0 sec
Base Flow:
***** LPM
In Hold:
0.0 sec
Compliance:
***** ml/cmH2O
I:E Ratio:
1.2:1
O2:
18.5 %
PIP:
1.9 cmH2O
Baro Press:
645.1 mmHg
MAP:
0.0 cmH2O
Gas:
Air
IPP:
0.0 cmH2O
PEEP:
0.0 cmH2O
Final Report
Test Parameter:
Minute Vol
Starting Value:
28.686 L
Starting Time:
10:45:30
Ending Time:
10:45:37
Elapsed Time:
0:00:07
% Limit:
5 %
# of Incidents:
1
# of Tests: 1