Fluke Biomedical VT Plus HF User Manual

Page 132

Advertising
background image

VT Plus HF

Operators Manual

A-6

O

u

tlet

Inlet

V

entilator

Inspiratory Hose

Expiratory Hose

Analyzer

TEST
LU

N

G

fec002.eps

Figure A-1. Circuit Connection for the Analyzer in Bi-directional Mode

Note

When using the Analyzer with a ventilator and a test lung, the bidirectional

breath detect mode is preferred. Only use the inspiratory and expiratory

breath detect modes if measurements cannot be made in the bi-directional

mode.

Additional breath detect modes were implemented for users that are accustomed to

performing measurements with flow devices that can only measure flow in one direction.

Base Flow

In the inspiratory and expiratory limb of the breathing circuit, some ventilators may have

a bias flow when no flow is going to or from the patient (test lung). This is implemented

differently on different ventilators and is called by different names by the various

ventilator manufactures. It may be called base flow, bias flow, flow-by, blow-by, or many

other names.
In the inspiratory- or expiratory-only breath detection modes, the Analyzer can measure

the base flow present. The detection and calculation of base flow may take 30 to 60

seconds. In expiratory-only mode, detection of base flow requires a sufficient end-

expiratory period and sufficient change in pressure. The use of lower-lung compliance

settings improves performance in expiratory mode.
During expiratory-only mode, the base flow is subtracted from the flow signal for

obtaining the ventilator parameters, such as tidal volume and peak flow. During

inspiratory-only mode, the base flow is not subtracted, because the base flow is typically

turned of during the inspiratory phase on most ventilators.

Advertising