Operation

Screen Objects

3

3-7

Title Bar

Each screen includes a title bar that shows the screen or mode name on the left. A number

indicating the screen number is shown in a small box on the right of the title bar:

fec039.bmp

9

Indicates the screen number

Data Plots

Except the

FULL

screen, each screen includes a data plot that shows a measured signal

(flow, pressure volume, or O

2

) vs. time:

fec045.bmp

New data is added to the plot starting on the left and progressing to the right. When the

plot has progressed to the right edge, the plot restarts on the left side of the plot. One line

of data from the previous plot is removed ahead of the new plot.
When a new screen is selected, the plot is displayed using a default plot range. Pressing

the

CLEAR

soft key restarts the plot and clears all previously plotted data and calculated

statistics. Pressing the

RESCALE

key changes the window of the data plot to a more

appropriate range based on the minimum and maximum values in the Statistics area.

Note

After pressing

CLEAR

, allow one full cycle to pass before pressing

RESCALE

.

Breath Parameters

Every screen can display up to four breath parameters:

fec047.bmp

These parameters are displayed only when breath detection is enabled. See Special

Functions for instructions on how to enable breath detection. Parameters are updated

following every breath. Units of measure for breath parameters are fixed. See

Introduction: Specifications for a listing of breath parameters and units. See Selecting

Breath Parameters for instructions on selecting which parameters are displayed for each

of the modes.