Connection and Setup

Setting Analyzer Operating Parameters

2

2-13

Setting Analyzer Operating Parameters

To adjust system information:
1. From the

Setup

screen, highlight

System

and press the

ENTER

soft key. The

System

menu displays:

fec032.bmp

2. Change settings to pre-selected options or to numeric values by highlighting the

option and pressing the

MODIFY

soft key.

For example, if one selects

Filtering

, then presses

MODIFY

, he can scroll through

the available Filtering options. If one selects an option that requires a numeric entry,

such as

Time

, and presses

MODIFY

, the entered data is cleared, and the field can

accept numeric input from the keypad until the

ENTER

key is pressed.

Note

The option displayed when exiting the settings screen is saved and used for

future measurements.

The following is a description of how to change each system setting. All of these options

are accessed from the

System

menu of the

Setup

screen.

Time

The Analyzer displays hours, minutes, and seconds on the display. However, the user can

set only hours and minutes.
To set the time:
1. Highlight

Time

.

2. Press the

MODIFY

soft key to display the value field.

3. Input the desired value from the keypad. Enter time in the HHMM format where HH

is the hours in 25-hour format (0 to 23) and MM is the minutes, a number between 0

and 59.

Note

For hour and minute values less than 10, enter a leading zero. For

example, if the time is 9:05, enter 0905.